Appendix B: Performance Verification
B-34
AWG610 Arbitrary Waveform Generator User Manual
Pulse Response Tests
This procedure checks the pulse response characteristics of the AWG610
Arbitrary Waveform Generator output waveforms at amplitudes of 1 V.
Equipment
required
A 50ĂW SMAĂcoaxial cable and an oscilloscope (TDS820).
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use the 50 W SMA coaxial cable to connect the AWG610 Arbitrary
Waveform Generator CH1 output connector to the oscilloscope CH1 input
connector (see Figure B–12).

AWG610 Arbitrary Waveform Generator
Oscilloscope (TDS820)
50 W SMAĂcoaxial cable
Figure B-12: Pulse response initial test hookup
2. Set the oscilloscope controls as follows:
Vertical . ........................ CH1
CH1 coupling . ............... DCifapplicable
CH1 scale . ................. 250mV/div (200 mV for TDS820)
Horizontal
Sweep . .................... 500ps/div
Trigger
Source . .................... CH1
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto