Vista de pagina 69
Command Descriptions
3–46
Option 01 VXI Interface Module User Manual
Command Response
trigger:sequence:
source immediate
Define a trigger source of “immediate”. This means to
sequence through the scan list without waiting for a trigger
event.
trigger:sequence:count
5
Sequence through the entire scan list five times.
route:close:dwell
gp,.5
Wait 0.5 seconds after closing a relay on the VX4350.
initiate:immediate Begin sequencing through the scan list.
*OPC Set the Operation Complete bit of the Standard Event Status
register after sequencing through the scan list five times.