Tektronix Thermometer 070-9472-02 Manual de usuario Pagina 26

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 25
Operating Basics
14
P5205 Instructi on Manual
With extension leads
Without extension leads
50 ns/div 50 ns/div
250 mv/div
250 mv/div
Figure 4: HF Transient response with and without extension leads
Probe Loading
When you touch your probe tip to a circuit element, you are
introduci ng a ne w resistance, capacitance, and inductance into the
circuit.
Frequency and impedance of the source dete rmine how much the
probe loads the circui t you are measuring. As the frequency of the
source starts to increase beyond 1 kHz, the input impe dance of the
probe begins to decrease. The lower the impeda nce of the probe
relative to that of the source, the more the probe loa ds t he circuit
under test. For a graph of frequency versus input impedance, re fer to
Figure 6 on page 19.
The probe has virtually no loading effect on sources with relatively
low impedance and low frequency.