Tektronix Thermometer P6249 Manual de usuario Pagina 39

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 82
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 38
P6249 4 GHz 5X Active Probe Instruction Manual
27
Specifications
These specifications apply to a P6249 probe when used with a
TDS 694C oscilloscope.
The probe and oscilloscope must first be allowed to warm up for
20 minutes before measurements are taken.
CAUTION. Do not apply voltages beyond the non-destructive input
voltage range to the probe. Damage to the probe or circuit under test
may result.
Table 2: Warranted electrical specifications
DC Attenuation Accuracy (probe only) 0.2 ±2% (excludes offset error)
Output Zero ±10 mV or less at output of probe
Rise Time (probe only) 120 ps
Table 3: Typical electrical characteristics
Bandwidth, (probe only) 4 GHz (See Figure 9)
Linear Input Dynamic Range 2.0 V to +2.0 V.
(Equivalent to 0.4 V to +0.4 V
at the output of the probe.)
Linearity
"0.1% over a dynamic range of 1.75 V to
+1.75 V
"1.0% over a dynamic range of 2.0 V to
+2.0 V
Nondestructive Input Voltage Range
30 V to +30 V (DC + peak AC)
Input Resistance
20 kW at DC. (See Figure 8)
Input Capacitance 1.0 pF
Vista de pagina 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 81 82

Comentarios a estos manuales

Sin comentarios