Performance Verification
VITS100 NTSC VITS Inserter Instruction Manual
5–21
2. Set the TSG/SPG Multiburst output as shown in Table 5–4.
Table 5–4: Multiburst Controls
Control
Setting
Frequency Range High
Markers On
Composite On
Amplitude Full
Sweep On
3. Connect 75 W coax from detector head output to peak-to-peak detector +
input.
4. Enable the peak-to-peak detector + input.
5. Set + input level to light the green LED (full clockwise).
6. Connect 75 W cable from the peak-to-peak detector output to the test
oscilloscope input.
7. Measure the sweep envelope. See Figure 5–11.
100
90
10
0%
Figure 5–11: Frequency response display
8. Remove the BNC female to-female adaptor.