Specifications
5–10
1780R-Series Operators Manual
Table 5–4: Waveform Monitor Horizontal Deflection System (Cont.)
Characteristics
Supplemental
Information
Performance
Requirements
RGB/YRGB
Staircase Input
Amplitude
A +10 V input will result in a
horizontal display of 9 divisions
±1.4 major divisions.
Ground to +10 V. +10 V
corresponds to left side of
the CRT.
Staircase Operating
Signal
DC Signal levels plus peak AC,
not to exceed –12 to +12 volts.
Max AC signal volts 12 V peak-to-peak. Field or line rate sweeps.
Sweep Length
RGB 2 Field = 27 – 33% of nor-
mal. 1 Line = 27 – 33% of
normal. 2 Line = 27 – 33%
of normal.
YRGB 2 Field = 20 – 25% of nor-
mal. 1 Line = 20 – 25% of
normal. 2 Line = 20 – 25%
of normal.
Sweep Repetition
Rate
Field or line rate of displayed
video or external sync signal as
selected by front-panel HOR-
IZONTAL controls.
Requires 1H or 1Fld sweep
selection.
External Horizontal Input Used for ICPM measure-
ments.
Sensitivity Direct coupled 0 – 5 V.
Sawtooth input of up to 5 V
is nominally a 10 division
horizontal sweep.
Input Impedance
10 kW.
Specifications
5–10
1780R-Series Operators Manual
Table 5–4: Waveform Monitor Horizontal Deflection System (Cont.)
Characteristics
Supplemental
Information
Performance
Requirements
RGB/YRGB
Staircase Input
Amplitude
A +10 V input will result in a
horizontal display of 9 divisions
±1.4 major divisions.
Ground to +10 V. +10 V
corresponds to left side of
the CRT.
Staircase Operating
Signal
DC Signal levels plus peak AC,
not to exceed –12 to +12 volts.
Max AC signal volts 12 V peak-to-peak. Field or line rate sweeps.
Sweep Length
RGB 2 Field = 27 – 33% of nor-
mal. 1 Line = 27 – 33% of
normal. 2 Line = 27 – 33%
of normal.
YRGB 2 Field = 20 – 25% of nor-
mal. 1 Line = 20 – 25% of
normal. 2 Line = 20 – 25%
of normal.
Sweep Repetition
Rate
Field or line rate of displayed
video or external sync signal as
selected by front-panel HOR-
IZONTAL controls.
Requires 1H or 1Fld sweep
selection.
External Horizontal Input Used for ICPM measure-
ments.
Sensitivity Direct coupled 0 – 5 V.
Sawtooth input of up to 5 V
is nominally a 10 division
horizontal sweep.
Input Impedance
10 kW.
Vista de pagina 161
1 2 ... 157 158 159 160 161 162 163 164 165 166 167 ... 187 188

Comentarios a estos manuales

Sin comentarios

Jenway Genova Plus French manuels

Manuels d'utilisation et guides de l'utilisateur pour Équipement Jenway Genova Plus French.
Nous fournissons des manuels en pdf 1 Jenway Genova Plus French à télécharger gratuitement par type de document : Manuel de l'utilisateur






Plus de produits et de manuels pour Équipement Jenway

Modèles Type de document
Genova Plus German Manuel de l'utilisateur   Jenway Genova Plus German Benutzerhandbuch, 116 pages
6715 Manuel de l'utilisateur   Jenway 6715 User Manual, 181 pages