Tektronix CSA8000B Manual de usuario Pagina 157

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 354
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 156
Measuring Waveforms
CSA8000B & TDS8000B User Manual
3-83
Use the procedure that follows to set gates on a measurement source, which
forces the measurement to be taken over a segment of the waveform (otherwise,
the entire waveform feeds the measurement).
Overview To gate a measurem ent Related control elements and resources
Prerequisites 1. Set up as f rom last procedure.
See To Take an Automatic Measurement on
page 3--24
Access the
gates
2. From the application menu bar, select Setup, and then
select Measurement. See right.
Enable and
position the
gates
3. Select the Region tab to expose the gate controls. Click
to check the box as indicated at right to turn gating on
and to display t he gates on screen.
4. If Annotations are not on, click the Annotations box, or
the gates wil l not display.
5. Use the G1 (Gate1) and G2 spin controls (or click ands
type in values--see right) to adjust the gates on screen
such that the area to measure i s bet ween the gates.
Tip. Values are entered as a % of the waveform,
displayed from left to right. If no keyboard is installed,
access the virt ual keyboard and use the t ouch screen to
enter values.
Check to display gates
Vary to position gates
A
ccess to
v
i
r
tual keyboa
r
d
To Localize a
Measurement
Vista de pagina 156
1 2 ... 152 153 154 155 156 157 158 159 160 161 162 ... 353 354

Comentarios a estos manuales

Sin comentarios