Using Masks, Histograms, and Waveform Databases
3- 152
CSA8000B & TDS8000B User Manual
Masks are created by connecting the points independently of the order they are
entered. Points are connected by sorting the points in left-to-right order and
grouping them across a diagonal from the left-most point to the right-most point.
If two points share a horizontal position along either the left or right edge of the
mask, the diagonal runs from the top left-most point to the bottom right-most
point. Points below the diagonal form the bottom boundary of the mask; points
above it form the top boundary. Use this procedure to create a new mask:
Overview To create a new mask Related control elements & resources
Prerequisites 1. The instrument must have at least one waveform turned
on and the Mask Set up di alog box displ ayed.
See Displaying Waveforms on page 3--53 for
information on displaying waveforms.
Select and di splay
a user-defined
mask
2. To create a mask from scratch, select User in the Comm
Standard s election li s t.
To Create a New Mask