Performance Tests
Performance Verification Procedures
75 Cable
75
Terminator
PAL Signal
Source
TSG121
COMPST
Figure 1-36: Line Count Accuracy Test Hookup
Press the main-menu Trigger Position.
Press the side-menu to Set to 50%.
Press the main-menu Horiz Pos.
Press the side-menu to Set to 50%.
Use the HORIZONTAL POSITION knob to move the falling edge
of the sync pulse to two divisions to the left of center screen (See
Figure 1-37).
Figure 1-37: Line Count Accuracy Test Setup Waveform
Press CURSOR.
Press the main-menu Function.