Tektronix TDS544A Manual de usuario Pagina 85

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 114
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 84
Performance Tests
TDS 520A, 524A, 540A, & 544A Performance Verification
c. Check Sync Duty Cycle
Connect the pulse generator (PG502) through a 50 cable and a
50 terminator to the oscilloscope CH1 input (See Figure 1-39).
50 Cable
Pulse
Generator
PG502
50
Terminator
Figure 1-39: PG502 Setup for Sync Duty Cycle Test
Turn the pulse generator OUTPUT (VOLTS) control until the
signal on the oscilloscope shows a one division negative going
pulse (See Figure 1-40).
NOTE
You may need to adjust the trigger level control to obtain a stable
trigger.
Vista de pagina 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 113 114

Comentarios a estos manuales

Sin comentarios