Setting SONETSDH Receive Parameters
3–110
CTS 710 SONET Test Set User Manual
Demapping a Tributary Signal (Add/Drop/Test Option Only)
To demap and test a tributary signal, you set Mapping to the
appropriate tributary mapping.
NOTE. The CTS 710 cannot demap a tributary signal and transmit a
tributary signal at the same time.
When you use tributary signal mapping, you must also specify the
virtual tributary under test and payload (see Figure 3–39).
Figure 3–39: Setting Tributary Mapping
To specify the VT Under Test and set the payload:
6. Specify the VT Under Test as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RECEIVE RECEIVE
SETTINGS
VT Under Test as appropriate
Setting SONETSDH Receive Parameters
3–110
CTS 710 SONET Test Set User Manual
Demapping a Tributary Signal (Add/Drop/Test Option Only)
To demap and test a tributary signal, you set Mapping to the
appropriate tributary mapping.
NOTE. The CTS 710 cannot demap a tributary signal and transmit a
tributary signal at the same time.
When you use tributary signal mapping, you must also specify the
virtual tributary under test and payload (see Figure 3–39).
Figure 3–39: Setting Tributary Mapping
To specify the VT Under Test and set the payload:
6. Specify the VT Under Test as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RECEIVE RECEIVE
SETTINGS
VT Under Test as appropriate