Tektronix 784A Manual de usuario Pagina 146

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 294
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 145
Performance Tests
4–70
TDS 684A, TDS 744A, & TDS 784A Service Manual
Positive pulse
Negative pulses
Figure 4–28: Triggered Signal Range Test – 300 mV
n. Use the keypad to set the fine scale to 75 mV (press 75, SHIFT, m, then
ENTER).
o. CONFIRM that the TRIG’D LED stays lighted and that the waveform
on screen is stable. That is, it does not move horizontally or vertically.
Also, CONFIRM that the waveform on the screen has one positive pulse
and a number of negative pulses. See Figure 4–29.
Vista de pagina 145
1 2 ... 141 142 143 144 145 146 147 148 149 150 151 ... 293 294

Comentarios a estos manuales

Sin comentarios