Tektronix 784A Manual de usuario Pagina 175

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 294
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 174
Adjustment Procedures
5–16
TDS 684A, TDS 744A, & TDS 784A Service Manual
a. Access Inner Probe Tip and Adjustment Ports:
H The probe tip should be exposed from the procedure Measure Probe
Bandwidth. If not, follow the instructions in Figure 5–5 to expose
the probe tip.
H Follow the instructions in Figure 5–7 to remove the probe body
covers.
Use a small standard screwdriver to
pry between the cover and metal cord
connector to pop off cover.
Repeat for lower cover.
Figure 5–7: Exposing the Probe Body
b. Hook up test-signal source:
H Connect the high-amplitude output of a calibration generator,
through a 50 W precision cable to the input of a tunnel diode pulser.
See Figure 5–8.
H Connect the output of the tunnel diode pulser through a 10X
attenuator to CH 1.
H Set the triggering level of the tunnel diode pulser to minimum.
Vista de pagina 174
1 2 ... 170 171 172 173 174 175 176 177 178 179 180 ... 293 294

Comentarios a estos manuales

Sin comentarios