Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-25
MEMory:STATe:RECall:AUTo
This command sets or queries whether to enable the automatic recall of last setup
memory when powered-on. The next time you apply the power, the arbitrary/func-
tion generator will automatically recall the settings you used when you powered off
the instrument.
If you select OFF, the default setups are recalled when you power on the instru-
ment.
Group
Memory
Syntax
MEMory:STATe:RECall:AUTo {ON|OFF|<NR1>}
MEMory:STATe:RECall:AUTo?
Arguments
ON or <NR1>0 enables the recall of the setup memory you last used before
powering off the instrument.
OFF or <NR1>=0 disables the last setup recall function.
Returns
<NR1>
Examples
MEMory:STATe:RECall:AUTo ON
sets the instrument to recall the last setup memory when powered-on.