Tektronix Thermometer P7240 Manual de usuario Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 46
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
P7240 4 GHz 5X Active Probe User Manual
13
Functional Check
A functional check may be performed using the PROBE COM-
PENSATION connections on the front panel of the oscilloscope. See
Figure 3.
Figure 3: Probe functional check connections
1. Connect the probe to a TekConnect receptacle on the oscillo-
scope.
2. Set the oscilloscope to display the appropriate probe channel.
3. Using a ground lea d and SMT KlipChip, connect the probe
ground to the outer ri ng of the PROBE COMPENSATION
connector on the oscilloscope.
4. Using a standard t ip, hold the probe to the center conductor of the
PROBE COMPENSATION BNC connector on the oscilloscope.
5. Press AUTOSET, or adjust the oscilloscope to display a stable
calibration waveform. (You may need to adjust the probe offset to
display the waveform.)
Vista de pagina 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 45 46

Comentarios a estos manuales

Sin comentarios