Tektronix AWG710B Manual de usuario Pagina 281

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 580
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 280
The APPL Menu
AWG710&AWG710B Arbitrary Waveform Generator User Manual 3-151
Input data
The specified pattern or waveform file is used as input data. When a pattern data
file is specified for input, the application reads only the MSB bits (DATA7). When
a waveform file is specified, this process converts the values equal to or greater
than 0.5 to a logic 1, and the values less than 0.5 to a logic 0.
The Pre–defined patterns shown in Table 3-41 are incorporated in the application:
Table 3-41: Pre–defined patterns
Pattern items Descriptions
PN9 9–bits M–series pseudo random pulse.
PN15 15–bits M–series pseudo random pulse.
100100
10001000
1000010000
1010101010
100000100000
1000000010000000