Performance Verification
4–10
THS710A, THS720A, THS730A & THS720P Service Manual
This test checks the bandwidth of channel 2.
1. First check the channel 1 bandwidth using the previous test. Then, perform
these additional steps to check the channel 2 bandwidth:
SCO
PE
CH 1
WAVEFORM
OFF
CH 2
TRIGGER Trigger Source Ch2
ME
AS
Gating & High-Low
Setup
High-Low Method Min/Max
Select Measrmnt Pk-Pk*
OK Select
Measrmnt
* You may need to press Select Page to display this selection.
2. Connect the TekScope instrument to the leveled sine wave generator as
shown below.
Leveled
sine wave
generator
Ch 2
Output
50 feedthrough
terminator
BNC cable
Check Oscilloscope
Channel 2 Bandwidth