Commands Listed in Alphabetical Order
BUS:B<x>:SPI
:FRAMING
Sets or returns the type of SPI framing.
Group
Bus
Syntax
BUS:B<x>:SPI:FRAMING {SS|IDLEtime}
BUS:B<x>:SPI:FRAMING?
Arguments
SS species framing by SS (non 2-wire).
IDLEtime speciesframingbyIdleTime(2-wire).
Examples
BUS:B1:SPI:FRAMING SS sets the SPI framing type to SS.
BUS:B1:SPI:FRAMING IDELTIME might return BUS:B1:SPI:FRAMING
IDELTIME
indicating the SPI framing type is set to IDLETIME.
BUS:B<x>:SPI:IDLETime
Sets or returns the SPI bus idle time in seconds for the specied bus.
Group
Bus
Syntax
BUS:B<x>:SPI:IDLETime <NR3>
BUS:B<x>:SPI:IDLETime?
Arguments
NR3 is the SPI bus idle time in seconds.
Examples
BUS:B1:SPI:IDLETIME 100.0000E-9 sets the idle time to 100 ns.
BUS:B1:SPI:IDLETIME? might
return
BUS:B1:SPI:IDLETIME
100.0000E-9
indicating the idle time is set to 100 ns.
BUS:B<x>:SPI{:SELect|:SS}:POLARity
Sets or returns the SPI SS polarity for bus <x>, where x is the bus number.
Conditions
This command requires a DPO4EMBD application module.