Commands Listed in Alphabetical Order
TRIGger:A:LO
GIc:INPut? (Query Only)
Returns the logic input values for all channels. If a clock channel is dened,
it returns the clock source and edge.
Group
Trigger
Syntax
TRIGger:A:LOGIc:INPut?
Examples
TRIGGER:A:LOGIC:INPUT? might return :TRIGGER:A:LOGIC:INPUT:CH1
HIGH;CH2 X;CH3 X
indicating that a logic high is expected on channel 1 while
channel 2 and channel three are “don’t care.”
TRIGger:A:LOGIc:INPut:CH<x>
Sets or returns the logical input condition for the channel specied by <x>.
Group
Trigger
Syntax
TRIGger:A:LOGIc:INPut:CH< x> {HIGH|LOW|X}
TRIGger:A:LOGIc:INPut:CH< x>?
Arguments
HIGH species the logic high.
LOW species the logic low.
X species a "don’t care" state.
Examples
TRIGGER:A:LOGIC:INPUT:CH1 ? might return :TRIGGER: LOGIC:INPUT:CH1
X
indicating that the setting for the A logic trigger input to channel 1 does not
matter.
TRIGGER:A:LOGIC:INPUT:CH2 HIGH sets the A logic trigger input to logic
HIGH for channel 2.
TRIGger:A:LOGIc:INPut:CLOCk:EDGE
Sets the polarity of the clock channel.
Group
Trigger
2-432 MSO4000 and DPO4000 Series Programmer Manual