Vista de pagina 199
Commands Listed in Alphabetical Order
Syntax
HORizontal:RE
COrdlength <NR1>
HORizontal:RECOrdlength?
Arguments
<NR1> represents the supported values for horizontal record lengths, which are:
1000, 10000, 100000, 1000000, or 10000000.
Examples
HORIZONTAL:RECORDLENGTH 10000 species that 10000 data points will be
acquired for each record.
HORIZONTAL:RECORDLENGTH? might return : HORIZONTAL:RECOrdlength
1000
indicating that the horizontal record length is equal to 1000 d ata points.
HORizontal:SAMPLERate
Sets or returns the current horizontal sample rate.
Group
Horizontal
Syntax
HORizontal:SAMPLERate <NR3>
HORizontal:SAMPLERate?
Arguments
<NR3> is the rate in samples per second.
HORizontal:SCAle
Sets or returns the time base horizontal scale.
Group
Horizontal
Syntax
HORizontal:SCAle <NR3>
HORizontal:SCAle?
Arguments
<NR3> species the range from 400 ps to 1000 s, depending on the record length.
Examples
HORIZONTAL:SCALE 2E-6 sets the main scale to 2µs per division.
HORIZONTAL:SCALE? might return :HORIZONTAL:MAIN:SCA LE 2.0000E-06
indicating that the main scale is currently set to 2 µs per division.