Commands Listed in Alphabetical Order
DESkew:DISPl
ay
Sets or returns the state of the deskew table display.
Group
Vertical
Syntax
DESkew:DISPlay {OFF|ON|0|1}
DESkew:DISPlay?
Arguments
OFF or 0 turns off the deskew table d isplay.
ON or 1 turns on the deskew table display.
DIAg:LOOP:OPTion
Sets the self-test loop option.
Group
Calibration a nd Diagnostic
Syntax
DIAg:LOOP:OPTion {ALWAYS|FAIL|ONFAI L|ONCE|NTIMES}
Arguments
ALWAYS continues looping until the self tests (diagnostics) are stopped via the
front panel or by an oscilloscope command.
FAIL causes looping until the first self test (diagnostic) failure or until self tests
(diagnostics) are stopped.
ONFAIL causes looping on a spe cifi c test group as long as a FAIL status is
returned from the test.
ONCE executes se lf test (diagnostics test) sequence once.
NTIMES runs “n” number of loops.
Examples
DIAG:LOOP:OPTION ONCE runs one loop of self tests.
DIAg:LOOP:OPTion:NTIMes
Sets the self-test loop option to run N times.
Group
Calibration a nd Diagnostic
2-154 MSO4000 and DPO4000 Series Programmer Manual