Vista de pagina 476
Commands Listed in Alphabetical Order
Related Commands
TRIGger:B:EVE
NTS?
Arguments
<NR1> is the number of B trigger events, which can range from 1 to 10,000,000.
Examples
TRIGGER:B:EVENTS:COUNT 4 species that the B trigger will occur four trigger
events after the A trigger.
TRIGGER:B:EVENTS:COUNT? might return :TRIGGER:B:EVENTS: COUNT 2
indicating that two events must occur after the A trigger b efore the B trigger
can occur.
TRIGger:B:LEVel
Sets or returns the level for the B trigger.
Group
Trigger
Syntax
TRIGger:B:LEVel {TTL|<NR3>}
TRIGger:B:LEVel?
Related Commands
TRIGger:A:LEVel, TRIGger:B, TRIGger:B:EDGE:SOUrce
Arguments
TTL species a preset TTL high level of 1.4 V.
<NR3> is the B trigger level, in volts.
Examples
TRIGGER:B:LEVEL TTL sets the B trigger level to 1.4 V.
TRIGGER:B:LEVEL? might return :TRIGGER:B:LEVEL 173.0 000E-03
indicating that the B trigger level is currently set at 173 mV.
TRIGger:B:LEVel:CH<x>
Sets or returns the B trigger level for channel <x>, where x is the channel number.
Each Channel can have an independent Level.
Group
Trigger
Syntax
TRIGger:B:LEVel:CH<x> {ECL|TTL|<NR3 >}
TRIGger:B:LEVel:CH<x>?
MSO4000 and DPO4000 Series Programmer Manual 2-467