Commands Listed in Alphabetical Order
Arguments
<NR3> is the mid
voltage reference level in volts.
Examples
POWER:SWLOSS:REFLEVEL:ABS OLUTE:GATEMID 100 sets the absolute gate
mid reference voltage to 100.0000 V.
POWER:SWLOSS:REFLEVEL:ABS OLUTE:GATEMID ? might return
:POWER:SWLOSS:REFLEVEL:AB SOLUTE:GATEMID 50.0000 indicating that
the gate mid reference v oltage is set to 50.0000 V.
POWer:SWLoss:REFLevel:ABSolute:LOWCurrent
Sets or returns the low current reference level used in switching loss power
measurements in amperes.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:SWLoss:REFLevel:ABS olute:LOWCurrent <NR3>
POWer:SWLoss:REFLevel:ABS olute:LOWCurrent?
Arguments
<NR3> is the low voltage current level in amperes.
Examples
POWER:SWLOSS:REFLEVEL:ABS OLUTE:LOWCURRENT 2.8 sets the absolute
reference low current to 2.8000 A.
POWER:SWLOSS:REFLEVEL:ABS OLUTE:LOWCURRENT ? might return
:POWER:SWLOSS:REFLEVEL:AB SOLUTE:LOWCURRENT 5.0000 indicating that
the absolute reference low current is set to 5.0000 A.
POWer:SWLoss:REFLevel:ABSolute:LOWVoltage
Sets or returns the low voltage reference level used in switching loss power
measurements in volts.
Conditions
This command requires a DPO4PWR application module.
Group
Power
2-304 MSO4000 and DPO4000 Series Programmer Manual