Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASUrement:STATIstics:MOD E {OFF|ON}
MEASUrement:STATIstics:MOD E?
Related Commands
MEASUrement:STATIstics
Arguments
OFF turns all measurements off. This is the default value.
ON turns on statistics and displays all statistics for each measurement.
Examples
MEASUREMENT:STATISTICS:MOD E OFF turns statistic measurements off.
MEASUREMENT:STATISTICS:MOD E? might return
:MEASUREMENT:STATISTICS:MO DE ON indicating that measurement statistics
are turned on and all statistics are being displayed for each measurement.
MEASUrement:STATIstics:WEIghting
Sets or returns the time constant for mean and standard deviation statistical
accumulations.
Group
Measurement
Syntax
MEASUrement:STATIstics:WEI ghting <NR1>
MEASUrement:STATIstics:WEI ghting?
Related Commands
MEASUrement:STATIstics:MODE
Arguments
<NR1> is the number of samples used for the mean and standard deviation
statistical accumulations.
Examples
MEASUREMENT:STATISTICS:WEI GHTING 4 sets statistical weighting to four
samples.
MEASUREMENT:STATISTICS:WEI GHTING? might return
:MEASUREMENT:STATISTICS:WE IGHTING 4 indicating that
measurement statistics weightingiscurrentlysetto4samples.
MSO4000 and DPO4000 Series Programmer Manual 2-235